Miniaturized cost-effective broadband spectrometer employing a deconvolution reconstruction algorithm for resolution enhancement
 
Miniaturized cost-effective broadband spectrometer employing a deconvolution reconstruction algorithm for resolution enhancement ■
 
Artem Shcheglov, Artem Shcheglov, Artem Shcheglov, Yunfeng Nie, Yunfeng Nie, Yunfeng Nie, Colas Schretter, Colas Schretter, Colas Schretter, Rob Heeman, Rob Heeman, Rob Heeman, Arthur van der Put, Arthur van der Put, Arthur van der Put, Willem Hoving, Willem Hoving, Willem Hoving, Hugo Thienpont, Hugo Thienpont, Hugo Thienpont, Peter Schelkens, Peter Schelkens, Peter Schelkens, Prof. Dr. Ir. Heidi Ottevaere, Prof. Dr. Ir. Heidi Ottevaere, Prof. Dr. Ir. Heidi Ottevaere
 
Abstract ■

We demonstrate a miniaturized broadband spectrometer employing a reconstruction algorithm for resolution enhancement. We use an opto-digital co-design approach, by firstly designing an optical system with certain residual aberrations and then correcting these aberrations with a digital algorithm. The proposed optical design provides an optical resolution less than 1.7 nm in the VIS-channel (400-790 nm) and less than 3.4 nm in the NIR-channel (760-1520 nm). Tolerance analysis results show that the components are within a commercial class, ensuring a cost-efficient design. We build the prototype with a size of 37x30x26 mm(3) and demonstrate that by applying a restoration algorithm, the optical resolution can be further improved to less than 1.3 nm (VIS-channel) and less than 2.3 nm (NIR-channel). (C) 2022 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement