Salar Tayebi, Adrián Gutiérrez Ruiz, Adrián Gutiérrez, Adrián Gutiérrez Ruiz, Ikram Mohout, Evelien Smets, Robert Wise, Johan Stiens, Manu L N G Malbrain
 

Journal of Clinical Monitoring and Computing

Contribution To Journal

Abstract 

This review presents an overview of previously reported non-invasive intra-abdominal pressure (IAP) measurement techniques. Each section covers the basic physical principles and methodology of the various measurement techniques, the experimental results, and the advantages and disadvantages of each method. The most promising non-invasive methods for IAP measurement are microwave reflectometry and ultrasound assessment, in combination with an applied external force.